1 BIODATA NARASUMBER •Nama : Dr. Ir. Fatimah Zulfah Padmadinata, DEA. •TTL : Bandung, 15 Oktober •Jabatan : Peneliti di LIPI •Instansi : Pusat Penelitian Sistem Mutu & Teknologi Pengujian – LIPI (P2SMTP) •Pengalaman Kerja : a.Dosen ITB (1978-1982) b.Peneliti Tamu di Jepang dan Belanda c.Kepala Pustan LIPI (1998-2001) d.Kepala PSSM LIPI (2001-2005) e.Kepala P2SMTP LIPI (2005-2008) f.Deputi Bidang Jasa Ilmiah LIPI (2011-2014)

2 KEMAMPUAN LABORATORIUM PENGUJIAN ALAT KESEHATAN DI LIPI (P2SMTP & Lab lainnya) Oleh: Dr. Ir. Fatimah Zulfah Padmadinata, DEA Peneliti P2SMTP - LIPI Seminar Analisa dan Evaluasi Hasil Pemetaan Sarana Produksi Alat Kesehatan dan Laboratorium Uji Alat Kesehatan Jogjakarta, 29-31 Mei 2014

3 Medical Devices

4 Alat kesehatan adalah instrumen, apparatus, mesin dan/atau implan yang tidak mengandung obat yang digunakan untuk mencegah, mendiagnosis, menyembuhkan dan meringankan penyakit, merawat orang sakit, memulihkan kesehatan pada manusia, dan/atau membentuk struktur dan memperbaiki fungsi tubuh. PERATURAN MENTERI KESEHATAN REPUBLIK INDONESIA NOMOR 1190/MENKES/PER/VIII/2010

5 Mengapa perlu diuji? •Produk yang terkait dengan keselamatan, kesehatan, keamanan dan lingkungan (K3L) perlu ditetapkan standarnya (SNI) yang diberlakukan secara wajib (PP102 – Th.2000) •Produk Alat kesehatan  sangat terkait dengan K3L •Pemenuhan kesesuaian atas standar dilakukan melalui prosedur pengujian yang dipercaya  laboratorium terakreditasi •Pengujian dilakukan untuk memastikan produk memenuhi faktor keamanan, kualitas/keandalan dan performans/spesifikasi

6 Resiko Bahaya dari Alat Kesehatan 1.Bahaya mekanik 2.Resiko dari kebakaran atau ledakan 3.Tidak berfungsi dengan benar 4.Output yang berlebih atau kurang 5.Infeksi 6.Salah penggunaan 7.Resiko dari paparan arus listrik yg tdk dikehendaki

7 Bahaya dari Peralatan Medik Listrik Persyaratan untuk alat medik listrik, harus memiliki tingkat keamanan yang lebih tinggi, karena: • pasien dihubungkan dengan beberapa peralatan medis listrik (pada perawatan intensiv) • pasien kontak langsung dengan rangkaian elektronik (ECG monitoring) • kontak divais langsung ke dalam jaringan

8 Efek Fisiologi dari Arus Listrik • Elektrolisa: mengalirnya arus searah ke dalam tubuh manusia dapat menyebabkan perpindahan ion2 dari polaritas berlawanan ke arah yang berlawanan, yang apabila berlangsung dalam orde beberapa menit dapat mengakibatkan pembengkakan akibat elektrolisa • Terbakar: jika arus listrik mengalir melalui zat yang mempunyai hambatan listrik, akan terjadi panas. Efek tersebut dapat berakibat terbakarnya jaringan tubuh tergantung dari besarnya arus • Kejang otot: Jika ada stimulasi listrik mengenai motor saraf atau otot, maka otot akan terstimulasi, berkontraksi. Apabila keberlangsungan stimulasi cukup lama, maka akan terjadi kejang pada otot. • Napas terhenti: apabila kekejangan terjadi pada otot antar rusuk (intercostal muscels) yang menghantarkan pernapasan, maka pernapasan terhenti

9 • Jantung berhenti: apabila otot jantung mengalami kejang, maka proses pemompaan dari jantung berhenti • Fibrilasi Ventricular: apabila arus listrik masuk ke dalam jantung cukup untuk membangkitkan bagian dari otot jantung, maka perambatan normal dari aktivitas listrik dalam jantung akan terganggu. Apabila gangguan tersebut cukup besar detak jantung dapat mencapai 300 beat/menit, sehingga proses pemompaan di jantung terhenti yang mengakibatkan kematian. Ambang batas arus untuk fibrilasi ventrikular adalah sekitar 75 sampai 400 mA.

10

11 JENIS ALAT KESEHATAN BERDASARKAN PENGGUNAAN •diagnosa, pencegahan, monitoring, treatment atau mengurangi/meringankan penyakit •diagnosa, pencegahan, monitoring, treatment atau mengurangi/meringankan atau penyembuhan luka atau kelumpuhan •Investigasi, penggantian atau modifikasi dari bagian anatomi atau proses fisiologi •kontrol

12 PERATURAN MENTERI KESEHATAN REPUBLIK INDONESIA NOMOR 1189/MENKES/PER/VIII/2010 TENTANG PRODUKSI ALAT KESEHATAN DAN PERBEKALAN KESEHATAN RUMAH TANGGA •(1) Produk alat kesehatan dan PKRT yang beredar harus memenuhi standar dan/atau persyaratan mutu, keamanan, dan kemanfaatan. •(2) Standar dan/atau persyaratan mutu, keamanan, dan kemanfaatan sebagaimana dimaksud pada ayat (1) sesuai dengan Farmakope Indonesia atau Standar Nasional Indonesia (SNI) atau Pedoman Penilaian Alat Kesehatan dan PKRT atau standar lain yang ditetapkan oleh Menteri.

13 KLASIFIKASI ALAT KESEHATAN

14 KLASIFIKASI BERDASARKAN RESIKO

15 Classification scheme •Class III - highest risk –heart valves, monitor directly contacting the cardiovascular system, dressing with drug •Class IIa & IIb - medium risk –foetal monitor, baby incubator, respirator •Class I - low risk –simple wound dressing, wheelchair

16 Class I Devices Contoh: pembalut elastik, sarung tangan periksa, dan instrumen bedah hand held Kontrol secara umum cukup untuk memastikan keamanan dan efektivitas

17 Kontrol Umum Larangan terhadap pemalsuan atau penggunaan merk yang tidak benar Pernyataan persyaratan premarket Good Manufacturing Practices (GMPs) Labeling Registrasi fasilitas pabrik Daftar tipe divais Pemeliharaan rekaman Perbaikan, penggantian atau pengembalian uang

18 Class II Devices 1.. • Contoh: kursi roda listrik, infusion pumps, dan tirai bedah • Kontrol secara umum saja tidak cukup memadai untuk jaminan keamanan dan efektivitas. Diperlukan metoda lainnya untuk jaminan tsb. • Juga diperlukan kontrol khusus

19 Kontrol Khusus Standard performans Post-market surveillance Patient registries Development and dissemination of guidelines/guidances Design controls Recommendations and other appropriate actions Tracking requirements

20 Class III Devices Insufficient information exists to determine that general and special controls are sufficient to provide reasonable assurance of the safety and effectiveness of such devices Such devices are: Life sustaining or life supporting Substantial importance in preventing impairment of human health; or Present unreasonable risk of illness or injury

21 STANDAR YG DIGUNAKAN OLEH INDUSTRI PERALATAN MEDIS •Standard memegang peranan penting dalam industri sebagai pemelihara konsistensi, menjaga keamanan dan meningkatkan efisiensi •Standard dapat dikelompokkan ke dalam cara pengaplikasiannya, yaitu kelompok vertikal dan horizontal. Standard vertikal pengaplikasiannya sangat spesifik untuk produk tertentu, sedangkan standard horizontal dapat diaplikasikan untuk produk secara umum. Contoh utk standard horizontal adalah: ISO 13485 dan ISO 14971. Sedangkan standard vertikal sebagai contoh adalah ISO 7197:2006 (neosurgical implants) •ISO 13485:2003 ; "best practices" pedoman untuk produk dimanufaktur, distribusi dan ditelusuri •ISO 60601; berhubungan dengan electrical safety and electromagnetic compatibility (EMC) •ISO 14971; merupakan seperangkat proses formal yg berhubungan dengan masalah resiko dan fokus terhadap bagaimana perusahaan dapat memproduksi produk dengan lebih aman •ISO 10993; bio kompatibilitas, Biological Evaluation of Medical Devices

22 IEC 60601-2-x standards rule •Within its scope a part 2 takes priority over all part 1 standards. •70 in IEC, many more in ISO •Specific to a category of products. •Modifications to requirements in part 1 •Addition, •Replacement or •Exemption •Dated references to part 1 standards. IEC 60601-1-2: Electromagnetic Compatibility IEC 60601-1-3: Radiation Protection IEC 60601-1-6: Usability IEC 60601-1-8: Alarms IEC 60601-1-9: Environment Modular Structure of IEC 60601-Series IEC 60601-1-10: Physiological Closed-Loop Controllers IEC 60601-1-11: Home Healthcare Environment ISO 14971 Risk management IEC 60601-1-12: Emergency Services Environment** IEC 60601-1: General requirements Part 1: Generic Part 2: Particular Collateral Standards ** publication 2013

23 KEMAMPUAN LAB PENGUJIAN BERDASARKAN STANDAR IEC 60601 P2SMTP - LIPI

24 Ruang Lingkup dan Tujuan Standar IEC 60601-1 Standar ini berlaku untuk keselamatan peralatan medik listrik, dan menitik beratkannya pada keselamatan yang terdiri dari beberapa persyaratan mengenai kemampuan kerja yang dikaitkan dengan keselamatan. Tujuan dari standar ini untuk menetapkan persyaratan umum bagi keselamatan peralatan medik listrik dan sebagai dasar untuk persyaratan keselamatan pada standar2 khusus lainnya.

25 Hal-hal yang diuji oleh standar IEC 60601-1 1. Arus bocor 2. Kekuatan dielektrik dari isolasi, yang diuji dengan tegangan tinggi 3. Ground bond continuity; keandalan kawat tanah terhadap arus tinggi (25 Amp) 4. Sisa tegangan, setelah kabel dicabut dalam 1 detik dari sumber daya. 5. Accessible parts voltage; tegangan yang ada pada sembarang bagian yang dapat disentuh 6. Enerji tersimpan; enerji yang tersimpan pada bagian yang dapat disentuh 7. Arus yang digunakan oleh alat 8. Daya yang digunakan oleh alat

26 No.Jenis Pengujian atau sifat-sifat yang diukurSpesifikasi, Metode pengujian, Teknik yang digunakan Ketersediaan Alat General Requirements for Safety 1.ClassificationIEC 601-1 butir 5100 % 2.Identification, Marking and DocumentsIEC 601-1 butir 6100 % 3.Power InputIEC 601-1 butir 7100 % 4.Special Environmental ConditionIEC 601-1 butir 10100 % 5.Requirements related to classificationIEC 601-1 butir 14100 % 6.Limitation of voltage and/or energyIEC 601-1 butir 15100 % 7.Enclosure and protective coversIEC 601-1 butir 1630 % 8.SeparationIEC 601-1 butir 1775 % 9.Protective Earthing, Functional Earthing, and Potential Equalization IEC 601-1 butir 18100 % 10.Continuous Leakage Currents and Patient Auxiliary Currents IEC 601-1 butir 19100 % 11.Dielectric StrengthIEC 601-1 butir 20100 % 12.Mechanical StrengthIEC 601-1 butir 21100 % 13.Moving PartsIEC 601-1 butir 22100 % 14.Surfaces, corners and edgesIEC 601-1 butir 23100 % 15.Stability and TransportabilityIEC 601-1 butir 2480 % 16.Expelled PartsIEC 601-1 butir 25100 % 17.Suspended MassesIEC 601-1 butir 28100 %

27 18.EMCIEC 601-1 butir 3640 % 19.Locations and Basic RequirementsIEC 601-1 butir 37100 % 20.Marking, Accompanying DocumentsIEC 601-1 butir 38100 % 21.Common Requirements for Category AP and Category APG Equipment IEC 601-1 butir 390 % 22.Requirements and Test for Category AP Equipment, Parts and Components Thereof IEC 601-1 butir 40100 % 23.Requirements and Test for Category APG Equipment, Parts and Components Thereof IEC 601-1 butir 41100 % 24.Excessive TemperaturesIEC 601-1 butir 42100 % 25.Fire PreventionIEC 601-1 butir 43100 % 26.Overflow, Spillage, Leakage, Humidity, Ingress of Liquid, Cleaning, Sterilization and Desinfection IEC 601-1 butir 4450 % 27.Pressure Vessels and Parts Subject to Pressure IEC 601-1 butir 450 % 28.Human ErrorIEC 601-1 butir 46100 % 29.Interruption of power supplyIEC 601-1 butir 49100 % 30.Accuracy of operating dataIEC 601-1 butir 50100 % 31.Protection against hazardous outputIEC 601-1 butir 51100 % 32.Abnormal operation and fault conditionsIEC 601-1 butir 52100 % 33.Environmental testIEC 601-1 butir 53100 % 34.Constructional requirements : GeneralIEC 601-1 butir 54100 %

28 35.Enclosure and CoversIEC 601-1 butir 55100 % 36.Component and general assemblyIEC 601-1 butir 56100 % 37.Mains parts, component and layoutIEC 601-1 butir 5730 % 38.Protective earthing – Terminal and Connections IEC 601-1 butir 5890 % 39.Construction and layoutIEC 601-1 butir 59100 %

29 KEMAMPUAN PENGUJIAN INKUBATOR ElectricalInfant IncubatorConditions for application to ME EQUIPMENT or ME SYSTEMS IEC 60601-2-19:2009 Clause 201.4.1 Essential Performance of Infant IncubatorIEC 60601-2-19:2009 Clause 201.4.3.101 SINGLE FAULT CONDITION for ME EQUIPMENT IEC 60601-2-19:2009 Clause 201.4.7 Components of ME EQUIPMENTIEC 60601-2-19:2009 Clause 201.4.8 Power InputIEC 60601-2-19:2009 Clause 201.4.11 Humidity preconditioning treatmentIEC 60601-2-19:2009 Clause 201.5.7 Determination of APPLIED PARTS and ACCESSIBLE PARTS IEC 60601-2-19:2009 Clause 201.5.9 ACCESSIBLE PARTSIEC 60601-2-19:2009 Clause 201.5.9.2 Test fingerIEC 60601-2-19:2009 Clause 201.5.9.2.1 Test hookIEC 60601-2-19:2009 Clause 201.5.9.2.2 Durability of markingsIEC 60601-2-19:2009 Clause 201.7.1.3 Limitation of voltage, current or energyIEC 60601-2-19:2009 Clause 201.8.4 ME EQUIPMENT intended to be connected to a power source by a plug IEC 60601-2-19:2009 Clause 201.8.4.3 Internal capacitive circuitsIEC 60601-2-19:2009 Clause 201.8.4.4

30 ElectricalInfant IncubatorSeparation of partsIEC 60601-2-19:2009 Clause 201.8.5 Separation of Patient ConnectionsIEC 60601-2-19:2009 Clause 201.8.5.2 F-TYPE APPLIED PARTSIEC 60601-2-19:2009 Clause 201.8.5.2.1 TYPE B APPLIED PARTSIEC 60601-2-19:2009 Clause 201.8.5.2.2 MAXIMUM MAINS VOLTAGEIEC 60601-2-19:2009 Clause 201.8.5.3 WORKING VOLTAGEIEC 60601-2-19:2009 Clause 201.8.5.4 Protective earth terminalIEC 60601-2-19:2009 Clause 201.8.6.2 Impedance and current-carrying capabilityIEC 60601-2-19:2009 Clause 201.8.6.4 Class II ME equipmentIEC 60601-2-19:2009 Clause 201.8.6.9 LEAKAGE CURRENTS and PATIENT AUXILIARY CURRENTS IEC 60601-2-19:2009 Clause 201.8.7 Distance through solid insulation or use of thin sheet material IEC 60601-2-19:2009 Clause 201.8.8.2 Dielectric strengthIEC 60601-2-19:2009 Clause 201.8.8.3 Mechanical strength and resistance to heatIEC 60601-2-19:2009 Clause 201.8.8.4.1 Resistance to environmental stressIEC 60601-2-19:2009 Clause 201.8.8.4.2

31 ElectricalInfant IncubatorCREEPAGE DISTANCES and AIR CLEARANCES IEC 60601-2-19:2009 Clause 201.8.9 Spaces filled by insulating compoundIEC 60601-2-19:2009 Clause 201.8.9.3 GeneralIEC 60601-2-19:2009 Clause 201.8.9.3.1 Insulating compound forming solid insulation between conductive parts IEC 60601-2-19:2009 Clause 201.8.9.3.2 Insulating compound forming a cemented joint with other insulating parts IEC 60601-2-19:2009 Clause 201.8.9.3.3 Thermal cyclingIEC 60601-2-19:2009 Clause 201.8.9.3.4 Measurement of creepage distances and air clearances IEC 60601-2-19:2009 Clause 201.8.9.4 Connections between different parts of ME EQUIPMENT IEC 60601-2-19:2009 Clause 201.8.10.3 Limitation of operating voltagesIEC 60601-2-19:2009 Clause 201.8.10.4.1 Connection cordsIEC 60601-2-19:2009 Clause 201.8.10.4.2 Instability – overbalanceIEC 60601-2-19:2009 Clause 201.9.4.2 Instability in transport positionIEC 60601-2-19:2009 Clause 201.9.4.2.1 Instability excluding transportIEC 60601-2-19:2009 Clause 201.9.4.2.2

32 ElectricalInfant IncubatorInstability from horizontal and vertical forcesIEC 60601-2-19:2009 Clause 201.9.4.2.3 Instability in transportIEC 60601-2-19:2009 Clause 201.9.4.3.1 Instability excluding transportIEC 60601-2-19:2009 Clause 201.9.4.3.2 Grips and other handling devicesIEC 60601-2-19:2009 Clause 201.9.4.4 Audible acoustic energyIEC 60601-2-19:2009 Clause 201.9.6.2.1 Sound level within the COMPARTEMNTIEC 60601-2-19:2009 Clause 201.9.6.2.1.101 Audible alarm sound levelIEC 60601-2-19:2009 Clause 201.9.6.2.1.102 Audible alarm sound level within COMPARTMENT IEC 60601-2-19:2009 Clause 201.9.6.2.1.103 Static forces due to loading from personsIEC 60601-2-19:2009 Clause 201.9.8.3.2 Dynamic forces due to loading from personsIEC 60601-2-19:2009 Clause 201.9.8.3.3 BarriersIEC 60601-2-19:2009 Clause 201.9.8.3.101 MATTRESS TRAYIEC 60601-2-19:2009 Clause 201.9.8.3.102 Support and mounting brackets for accessoriesIEC 60601-2-19:2009 Clause 201.9.8.101 Maximum temperature during NORMAL USEIEC 60601-2-19:2009 Clause 201.11.1.1

33 ElectricalInfant IncubatorAPPLIED PARTS intended to supply heat to a PATIENT IEC 60601-2-19:2009 Clause 201.11.1.2.1 APPLIED PARTS not intended to supply heat to a PATIENT IEC 60601-2-19:2009 Clause 201.11.1.2.2 Strength and rigidity required to prevent fire in ME EQUIPMENT IEC 60601-2-19:2009 Clause 201.11.2.1 Overflow in ME EQUIPMENTIEC 60601-2-19:2009 Clause 201.11.6.2 Ingress of water or particulate matter into ME EQUIPMENT and ME SYSTEMS IEC 60601-2-19:2009 Clause 201.11.6.5 Stability of Incubator TemperatureIEC 60601-2-19:2009 Clause 201.12.1.101 Uniformity of Incubator TemperatureIEC 60601-2-19:2009 Clause 201.12.1.102 Accuracy of skin Temperature SensorIEC 60601-2-19:2009 Clause 201.12.1.103 Accuracy between skin temperature and control temperature IEC 60601-2-19:2009 Clause 201.12.1.104 Accuracy of Incubator Temperature IndicationIEC 60601-2-19:2009 Clause 201.12.1.105 Accuracy of Incubator Temperature ControlIEC 60601-2-19:2009 Clause 201.12.1.106 Warm-up TimeIEC 60601-2-19:2009 Clause 201.12.1.107 Overshoot of Incubator TemperatureIEC 60601-2-19:2009 Clause 201.12.1.108

34 ElectricalInfant IncubatorAccuracy of Indication of Relative HumidityIEC 60601-2-19:2009 Clause 201.12.1.109 Oxygen ControlIEC 60601-2-19:2009 Clause 201.12.1.110 Air VelocityIEC 60601-2-19:2009 Clause 201.12.1.111 Weighing ScaleIEC 60601-2-19:2009 Clause 201.12.1.112 Alarm systemsIEC 60601-2-19:2009 Clause 201.12.3 Air circulation fanIEC 60601-2-19:2009 Clause 201.12.3.101 Exceeding LEAKAGE CURRENT or voltage limits IEC 60601-2-19:2009 Clause 201.13.1.3 Overheating of transformers in ME EQUIPMENT IEC 60601-2-19:2009 Clause 201.13.2.3 Failure of THERMOSTATSIEC 60601-2-19:2009 Clause 201.13.2.4 Failure of temperature limiting devicesIEC 60601-2-19:2009 Clause 201.13.2.5 Impairment of cooling that could result in a HAZARD IEC 60601-2-19:2009 Clause 201.13.2.7 Locking of moving partsIEC 60601-2-19:2009 Clause 201.13.2.8 Interruption and short circuiting of motor capacitors) IEC 60601-2-19:2009 Clause 201.13.2.9

35 ElectricalInfant IncubatorAdditional test criteria for motor operated ME EQUIPMENT IEC 60601-2-19:2009 Clause 201.13.2.10 Failures of components in ME EQUIPMENT used in conjunction with OXYGEN RICH ENVIRONMENTS IEC 60601-2-19:2009 Clause 201.13.2.11 General overload test conditionsIEC 60601-2-19:2009 Clause 201.13.2.13.1 ME EQUIPMENT with heating elementsIEC 60601-2-19:2009 Clause 201.13.2.13.2 ME EQUIPMENT with motorsIEC 60601-2-19:2009 Clause 201.13.2.13.3 ME EQUIPMENT RATED for non- CONTINUOUS OPERATION IEC 60601-2-19:2009 Clause 201.13.2.13.4 Mould stress relief testIEC 60601-2-19:2009 Clause 201.15.3.6 Fixing, prevention of maladjustmentIEC 60601-2-19:2009 Clause 201.15.4.6.1 Limitation of movementIEC 60601-2-19:2009 Clause 201.15.4.6.2 TransformersIEC 60601-2-19:2009 Clause 201.15.5.1.1 Dielectric strengthIEC 60601-2-19:2009 Clause 201.15.5.2 EARTH LEAKAGE CURRENT of MULTIPLE SOCKET-OUTLET IEC 60601-2-19:2009 Clause 201.16.6.2 PATIENT LEAKAGE CURRENTIEC 60601-2-19:2009 Clause 201.16.6.3

36 ElectricalInfant IncubatorElectromagnetic compatibility of ME EQUIPMENT and ME SYSTEMS IEC 60601-2-19:2009 Clause 201.17 Electromagnetic Compatibility – Requirement and Test See IEC 60601-1-2: 2007

37 Sistem uji inkubator bayi berdasarkan IEC 60601-1 dan IEC 60601-2-19

38 Sistem uji tekanan Sphygmomanometer dalam kondisi temperatur dan kelembaban ruang tertentu

39 Sistem dan metode uji refrigerator, vaccine or icepack freezer

40 Perancangan dan Pengembangan Sistem dan Metode Uji Pengaruh EMC (Elektromagnetic Compatibility) pada Kinerja Peralatan Kelistrikan dan Elektronika Berdasarkan Standar IEC 61000-4 Series

41 41 Keppres no. 103 tahun 2001

42 42 Kantor Pusat + 5 Puslit IPSK+ Oseanografi Puslit Fisika, Kimia, Metalurgi, KIM, SMTP Puslit Elkom, Info, Telimek, Geotek. UPT BPI,BIT UPT BPPTK UPT Karangsambung KR Purwodadi KR Bali UPT Mataram UPT Biota Laut Ambon UPT Biota Laut Tual UPT Biota Laut Biak UPT Biota Laut Bitung UPT Pengolahan Mineral Tg Bintang BB PTTG, Subang Puslit Biologi, Biotek, Limnologi, Pusbindiklat, UPT Biomaterial LOKASI SATKER LIPI (menyebar di seluruh Indonesia) UPT Loka Liwa Sts. Wamena

43 Bidang uji lainnya di Laboratorium LIPI

44

45

46

47

48

49